The Economical Series-2 Digital Force Gauges are developed as an entry gauge for basic force testing applications.
With its accuracy of ±0,5% and a 500Hz sampling rate and peak readings this device captures reliable readings for applications in both production and laboratory environments.
The compact Series-2 Digital Force Gauge comes in 8 force versions ranging from 10 to 2500N.
It features a backlit graphic display comes in an aluminum reversible housing and can be used as hand held or fixture mounted device.
Series-2 Force Gauges are compatible with a wide range of test stands and gripping fixtures.
The Digital Mark-10 Series Force Gauge range is available in a variation of 5 gauges with each gauge type available in a scale of force ranges.
Each type of gauge has its own specifications, the table below shows the main specifications of each gauge type.
All gauges types are equipped with a rechargeable battery, a back lit display and reversible aluminum housing.
The Series-# Digital Force Gauges are designed to be used for measuring tension and compression forces.
Mark-10 Digital Force Gauges are versatile instruments, in general they are combined with additional equipment to a purposely build system to perform the required tasks.
We are happy to assist you in finding the most suitable configuration for your needs. Please contact us for more information.
The Series-2 is supplied as a kit including:
Model | lbF | kgF | N |
---|---|---|---|
M2-2 | 2 x 0,002 | 1 x 0,001 | 10 x 0,01 |
M2-5 | 5 x 0,005 | 2.5 x 0,002 | 25 x 0,02 |
M2-10 | 10 x 0,01 | 5 x 0,005 | 50 x 0,05 |
M2-20 | 20 x 0,02 | 10 x 0,01 | 100 x 0,1 |
M2-50 | 50 x 0,05 | 25 x 0,02 | 250 x 0,2 |
M2-100 | 100 x 0,1 | 50 x 0,05 | 500 x 0,5 |
M2-200 | 200 x 0,2 | 100 x 0,1 | 1000 x 1 |
M2-500 | 500 x 0,5 | 250 x 0,2 | 2500 x 2 |
Gauge | M2-2 – M2-100 | M2-200 - M2-500 |
---|---|---|
Load Cell Thread | #10-32 UNF | 5/16-18 UNC |
Series-2 force gauge models M2-2 tot M2-500 all have the same dimensions
All dimensions are noted in Inches with metric sizes in (mm) behind.
Series-7 Professional | Series-5 Advanced | Series-4 Performance | Series-3 Basic | Series-2 Economical | |
---|---|---|---|---|---|
Accuracy | ±0,1% | ±0,1% | ±0,2% | ±0,3% | ±0,5% |
Sampling rate | 14.000 Hz | 7.000 Hz | 3.000 Hz | 2.000 Hz | 500 Hz |
Resolution | 1/5.000 | 1/5.000 | 1/2.000 | 1/1.000 | 1/1.000 |
Units of measurement | lbF, ozF, gF, kgF, N, kN, mN (depending on model) | lbF, ozF, gF, kgF, N, kN, mN (depending on model) | lbF, ozF, gF, kgF, N, kN, mN (depending on model) | lbF, gF, kgF, N (depending on model) | lbF, kgF, N |
Outputs | USB, RS-232, Mitutoyo, Analog | USB, RS-232, Mitutoyo, Analog | USB, RS-232, Mitutoyo, Analog | USB | USB |
- RS-232 | Configurable up to 230.400 baud | Configurable up to 115.200 baud | Configurable up to 115.200 baud | No | No |
- USB | Configurable up to 230.400 baud | Configurable up to 115.200 baud | Configurable up to 115.200 baud | Configurable up to 115.200 baud | Configurable up to 115.200 baud |
- Mitutoyo (Digimatic) | Serial BCD suitable for all Mitutoyo SPC-compatible devices | Serial BCD suitable for all Mitutoyo SPC-compatible devices | Serial BCD suitable for all Mitutoyo SPC-compatible devices | No | No |
- Analog | ±1 VDC, ±0,25% FS at capacity. Positive for compression, negative for tension | ±1 VDC, ±0,25% FS at capacity. Positive for compression, negative for tension | ±1 VDC, ±0,25% FS at capacity. Positive for compression, negative for tension | N/A | N/A |
- General purpose I/O | Three open drain outputs, one input | Three open drain outputs, one input | Three open drain outputs, one input | N/A | N/A |
- Set point | Three open drain lines | Three open drain lines | Three open drain lines | N/A | N/A |
Mesur-Lite software included | Yes | Yes | Yes | Yes | Yes |
Data output functionality | Automatic output or by pressing DATA button, or requesting from a PC or other device. | Automatic output or by pressing DATA button, or requesting from a PC or other device. | By pressing DATA button, or requesting from a PC or other device (currently displayed reading only). | By pressing DATA button, or requesting from a PC or other device (currently displayed reading only). | By pressing DATA button, or requesting from a PC or other device (currently displayed reading only). |
ASCII command set for external control | Full command set | Full command set | “?” only (request current reading) | “?” only (request current reading) | “?” only (request current reading) |
Interface w/MESUR® gauge and MESUR® Lite | Full communication | Full communication | Single point or continuous data output only | Single point or continuous data output only | Single point or continuous data output only |
Interface with ESM303 / 750 / 1500 test stands | Full communication, including PC control | Full communication, including PC control | Limited to force data, overload protection, and force set points | N/A | N/A |
Data memory | 5.000 readings, with output | 1.000 readings, with output | 50 readings, with output | No | No |
Data memory deletion | Any individual data point, or clear all | Any individual data point, or clear all | Any individual data point, or clear all | N/A | N/A |
Statistic calculations and output to PC | Max, min, mean, std. dev., number of readings, with output | Max, min, mean, std. dev., number of readings, with output | Max, min, mean, std. dev., number of readings, with output | N/A | N/A |
Break detection | Yes | No | No | No | No |
1st / 2nd peak detection | Yes | No | No | No | No |
Programmable footswitch command string | Yes | No | No | No | No |
Averaging mode | Yes | Yes | No | No | No |
External trigger mode | Yes | Yes | No | No | No |
Analog load bar graph shown on display | Yes | Yes | Yes | Yes | No |
Set points (for pass / fail indication) | Yes with high / low indicators, markers on analog bar graph, and outputs | Yes with high / low indicators, markers on analog bar graph, and outputs | Yes with high / low indicators, markers on analog bar graph, and outputs | Yes with high / low indicators, markers on analog bar graph, and outputs | No |
Password protection | Yes, for calibration and individual keys | Yes, for calibration and individual keys | No | No | No |
Peak readings always displayed | Yes | Yes | Yes | Yes | No (Peak mode must be selected) |
Programmable audible tones | Yes | Yes | Yes | Yes | No |
Calibration procedure | Configurable number of data points (up to 10 in each direction) | Configurable number of data points (up to 10 in each direction) | Configurable number of data points (up to 10 in each direction) | Configurable number of data points (up to 10 in each direction) | Configurable number of data points (up to 10 in each direction) |
Overload outputs | Dedicated pins for tension and compression | Dedicated pins for tension and compression | Dedicated pins for tension and compression | No | No |
Safe overloads | 200% of full scale (display shows warning at 110% and above) | 200% of full scale (display shows warning at 110% and above) | 200% of full scale (display shows warning at 110% and above) | 200% of full scale (display shows warning at 110% and above) | 200% of full scale (display shows warning at 110% and above) |
Accuracy | ±0,5% of full scale |
Sampling Rate | 500 Hz |
USB output | Configurable up to 115,200 baud |
Power | AC or rechargeable battery. Multi-step low battery indicator is displayed, gauge shuts off automatically when power is too low. |
Battery Life | Backlight on: up to 24 hours of continuous use Backlight off: up to 92 hours of continuous use |
Measurement Units | N, lbF, kgF |
Safe Overload | 200% of full scale (display shows “OVER” at 110% and above) |
Weight | M2-2 – M2-100: 0,33 kg [0.7 lb] M2-200 – M2-500: 0,41 kg [0.9 lb] |
Environmental Requirements | 0 - 40°C, max. 96% humidity, non-condensating |
Warranty | 3 years |
Country of Origin | USA |