Measuring range with standard probe with delay tip:
0.150 - 25.40mm
Total measuring range with P-E contact probe:
up to 915.0 mm
The TI-007X is a precision ultrasonic wall thickness gauge has been developed for high-resolution measurements on thin metals and plastics. The technology comes in an extruded aluminum housing, which has been sealed to resist even difficult environmental and production conditions.
The measuring range of the TI-007X wall thickness gauge is 0.15 - 25.4mm with standard probe T-402-5507. Using a single-element probe with delay tip, thin materials can be measured. The device then automatically removes any echo-echo (E-E) coatings and automatically switches to interface-echo mode (I-E) when measuring thicker materials. For the measurements of plastics, the user selects the plastic mode (PLAS) again with a graphite delay-tip.
In addition,the TI-007X offers the difference mode (DIFF) where the user sets a nominal value and the instrument calculates the ± difference between the nominal value and the measured wall thickness. In alarm mode, the user sets maximum and minimum values, and the LEDs on the front panel are triggered when the readings are outside the tolerance range.
The TI-007 can also be used with single element contact probes, extending the measurement range up to 923mm (depending on material and probe). Although the TI-007X does not have an internal memory, individual readings can be sent to a PC or other suitable data collector via the USB port. If a data store is needed, the TI-007DLX is the right alternative device choice.
The TI-007X wall thickness gauge comes in a complete set consisting of:
Everything delivered in a foam-lined plastic carrying case.
P/N | Description | Standard / Optional | Image |
---|---|---|---|
X-231-0000 | 1/4" x 3/8" for all metals (short) |
standard | |
X-241-0000 | 1/4" x 1/2" for all metals (long) |
standard | |
X-633-0000 | 1/4" x 3/8" Graphite-tip for thin plastics (short) |
optional |
The graphite tip is used in special plastic mode to measure plastics.
Minimum Radius for Convex Surfaces | 12.7 mm | |
Minimum Radius for Concave Surfaces | 63.5 mm | |
Minimum Headroom | 33.0 mm | |
Minimum Sample Diameter | 12.7 mm |
Maximum measuring range: | up to 924 mm (with contact P-E transducer) |
Measuring range with standard probe (T-402-5507): | 0.15 - 25.4 mm |
Resolution: | Switchable 0.001 mm or 0.01 mm |
Gain: | Automatic time-dependent amplification (TDG) with manual override |
Sound Velocity Range: | 305 - 18542 m/sec |
Probe (standard): | 1/4 "diameter, 15 MHz probe with delay-tip and cable (item T402-5507) |
Probe (optional): | Diverse single element probes with delay tip or contact probes (5 - 20 MHz) |
Cable: | 1.2m with Microdot / Lemo 00 connector |
Display: | Multifunction LCD display with 7 segments and backlight |
Backlight display: | Illumination selectable ON / OFF / AUTO and adjustable illumination intensity (Lo, Med, Hi) |
Display update rate: | 10 Hz (10 updates / second) |
Data output: | Output of individual measured values via USB (no internal memory). Corresponding model with data memory is the TI-007DLX. |
Operating temperature: | -30 - 75°C |
Battery type: | 2x AA 1.5V alkaline, 1.2V NiCad or 1.5 lithium AA cells (rechargeable batteries can be used) |
Battery duration: | Typically 35 hours with alkaline batteries and 18 hours with NiCad. |
Weight: | 308 grams |
Dimensions: | 64 x 131 x 32 mm |
Housing: | Extruded aluminum body with nickel-plated aluminum end caps (sealed with gasket) |
Class protection housing: | IP65 |
Certification: | NIST traceable and MIL-STD-45662A |
Key board: | Sealed membrane keypad |
Pulse Holwierde frequency: | 200 Hz (200 pulses / second) |
Warranty: | Measuring device: 5 years Test head: 90 days |