Temperature Range: | -32 to 535°C (-25 to 999°F) |
Accuracy: | (assumes ambient temperature of 23°C (73°F)) |
–32 to –26°C (-25 to -15°F): | ±3ºC (±5ºF) |
-26 to -18°C (-15 to 0°F): | ±2.5ºC (±4ºF) |
-18 to 23°C (0 to 73°F): | ±2ºC (±3ºF) |
23° to 510ºC (73º to 950ºF): | ±1% of reading or ±1ºC (±2ºF), whichever is greater |
Above 510ºC (950ºF): | ±1.5% of reading |
Repeatability: | ¡¾0.5% or ¡Â ¡¾1¨¬C (¡¾2¨¬F), whichever is greater |
Response Time: | ¡Ü 0.5 second (95% of reading) |
Spectral Response: | 8-14µm |
Emissivity: | Pre-set at 0.95 |
Ambient Operating Temp: | 0 to 50ºC (32 to 120ºF) |
Relative Humidity: | 10-90% |
Storage Temperature: | -20 to 60ºC (-13 to 158ºF) without battery |
Weight/Dimensions: | 360g (12oz) 205 x 160 x 55mm (8x6x2 in) |
Power: | 9V Alkaline or NiCd Battery |
Battery Life (Alkaline): | 4 hrs w/laser, flashlight & backlight on 20 hrs w/laser, flashlight & backlight off |
Laser (Class II): | SmartSight™ dual laser sighting system |
Typical Distance to Target: | 200 mm-600 mm (8 inches to 2 feet) |
Distance to Spot (D:S): | Optimized for use at 8 inches Approximately 16:1 at focus point |
Display Hold: | 7 seconds |
Temperature Display: | ºC or ºF selectable |
Display Resolution: | 0.2ºC (0.5ºF) |
Work area illumination: | Bright white LED (7100 millicandle) |
Warranty: | 1 year |