The TI-PVX Precision Ultrasonic Wall thickness gauge gives the user the necessary flexibility for both high resolution and penetration requirements.
It is the perfect tool for very precise and delicate measurements. This is due to the ability of the TI-PVX to use a variety of single element transducers for specific applications. These are: Standard Delay Line (acrylic and graphite tips for metals and thin plastics), Pencil Delay Line (tough access area on thin materials), and Contact transducers (variety of applications).
There are different viewing options available like RF waveform, +/- Rectified waveform, and Large Digits with Scan Bar.
Additional there is the time based B-Scan feature which displays a cross section of the test material and can display the profile of the opposite surface of the material.
Another available feature is the high speed scan function which speeds up the inspection process (32 measurements per second). With this feature the user can scan a larger area in a relatively short period of time. After the transducer is removed from the test material, the minimum thickness measured will appear on the display.
P/N | Description | Standard / Optional | Image |
---|---|---|---|
X-152-0000 | 1/16” Dia. x .450" L | Standard | |
X-752-0000 | 1/8" Dia. x .450" L | Optional |
P/N | Description | Standard / Optional | Image |
---|---|---|---|
X-231-0000 | 1/4" x 3/8" for all metals (Short) |
Standard | |
X-241-0000 | 1/4" x 1/2" for all metals (Long) |
Standard | |
X-633-0000 | 1/4" x 3/8" Graphite Tip for thin plastics (Graphite Short) |
Optional | |
X-532-0000 | 3/16" x 3/8" Conical Tip for all metals (Cone) |
Optional | |
X-533-0000 | 3/16" x 3/8" Graphite Tip for thin plastics (Graphite Cone) |
Optional |
The TI-PVX is supplied as a complete kit with :
Range: | Interface-Echo Mode: Steel 1.27mm - 25.4mm (0.05 - 1.0 in). Plastics from 0.127mm (.005in.). Echo-Echo Mode: Steel 0.15mm - 12.7mm (.006in -.500in). Pulse-Echo Contact Mode: Steel 1mm - 254mm (.040in - 10.0in). Plastics from 0.254mm (.010in). Echo-Echo Contact mode: Steel thru-paint 2.54mm - 76.2mm (.100in - 3.0in). |
Resolution: | (selectable) +/- 0.01 mm (.001in), or +/- 0.001mm (.0001in). |
Velocity Range: | .0492 to .3936 in./ms 1250 to 9999 meters/sec |
Units: | English & Metric |
Memory: | 16 megabit non-volatile ram. |
Data Logger: | 12.000 readings and waveforms (alpha numeric storage). |
Display: | 1/8 in. VGA grayscale display |
Setups: | 16 factory and 48 user-defined setups. |
Display Views: | A-Scan: Rectified +/- (half wave view) RF (full waveform view) B-Scan: Time based cross section view. Display speed of 15 secs per screen. Large Digits: Standard thickness view. Digit Height: 10mm (0.400 in). Scan Bar: 6 readings per second. Viewable in B-Scan and Large Digit views. Repeatability Bar Graph: Bar graph indicates stability of reading. |
Power Source: | Three 1.5V alkaline or 1.2V NiCad AA cells |
Battery Life: | Typically operates for 150 hours on alkaline and 100 hours on NiCad |
Power off: | Auto power off if idle 5 min. |
Operating Temperature: | -10° to 60°C (-14° to 140°F). |
Size: | 63.5 W x 165 H x 31.5 D mm (2.5 W x 6.5 H x 1.24 D inches). |
Weight: | 0.4 kg (0.9 lbs ) |
Warranty: | Gauge: 2 Years Probe: 90 Days |